Energy-dispersive X-ray spectrometry (EDX) (X-ray micro-analyser)

A photospectrometric method in which the sample under investigation is bombarded with an electron beam approximately 1 µm in diameter, so that the characteristic X-rays emitted from the irradiated area can then be quantitatively analysed using an X-ray spectrometer to determine the composition of the sample. What makes this method unique is that a very small, spatially confined section of the material can be analysed non-destructively. Consequently, the method is particularly well suited to the analysis of localised phenomena, such as the segregation of alloys or the diffusion layer of a solder joint. One disadvantage, however, is that whilst quantitative analysis is possible, it relates only to a single point on the sample and therefore requires various corrections. The accuracy is consequently lower than with X-ray fluorescence analysis. Furthermore, elements with an atomic number lower than that of magnesium can hardly be detected. By scanning the sample with the electron beam and varying the brightness of a Braun tube in proportion to the characteristic X-ray spectrum at each point, a continuous two-dimensional analysis (scan) is possible.